This document discusses defibrillation and defibrillators. It describes how defibrillation works to convert ventricular fibrillation into a normal heart rhythm through electrical shock. It discusses the need for and types of defibrillators, including internal defibrillators like implantable cardioverter defibrillators and external defibrillators like automated external defibrillators. It also covers defibrillator components like electrodes and how the devices analyze heart rhythms and deliver electrical shocks to the heart when needed.
1 of 20
Downloaded 66 times
More Related Content
2. defibrillators
1. 2/14/2014
1
DEFIBRILLATOR
BY
M. WASIM MUNIR
Bio-Instrumentation I
Defibrillation
2
ï‚—Defibrillation is based upon the understanding that contraction of the heart and the resulting circulation is under the control of the electrical conduction system of the heart
ï‚—Defibrillation is a common treatment for life- threatening cardiac dysrhythmias, ventricular fibrillation and ventricular tachycardia
ï‚—Defibrillation consists of delivering a therapeutic dose of electrical energy to the affected heart with a device called a defibrillator
Department of Biomedical Engineering, SSUET
2. 2/14/2014
2
NEED FOR A DEFIBRILLATOR
3
ï‚—Ventricular fibrillation is a serious cardiac emergency resulting from asynchronous contraction of the heart muscles
ï‚—Due to ventricular fibrillation, there is an irregular or rapid heart rhythm
Department of Biomedical Engineering, SSUET
NEED FOR A DEFIBRILLATOR
4
ï‚—Ventricular fibrillation can be converted into a more efficient rhythm by applying a high energy shock to the heart
ï‚—This sudden surge across the heart causes all muscle fibres to contract simultaneously
ï‚—Possibly, the fibres may then respond to normal physiological pacemaker pulses
ï‚—The instrument for administering the shock is called a DEFIBRILLATOR
Department of Biomedical Engineering, SSUET
3. 2/14/2014
3
5
Department of Biomedical Engineering, SSUET
TYPES OF DEFIBRILLATORS
Department of Biomedical Engineering, SSUET
6
Internal
External
4. 2/14/2014
4
TYPES OF DEFIBRILLATORS
a)Internal defibrillator
•Electrodes placed directly to the heart
•E.g. -Implantable CardioverterDefibrillator (ICD)
b)External defibrillator
•Electrodes placed directly on the heart
•E.g. -Automated External Defibrillator (AED)
7
Department of Biomedical Engineering, SSUET
DEFIBRILLATOR ELECTRODES
8
ï‚—Types of Defibrillator electrodes:-
a)Spoon shaped electrode
•Applied directly to the heart
b)Paddle type electrode
•Applied against the chest wall
c)Pad type electrode
•Applied directly on chest wall
Department of Biomedical Engineering, SSUET
5. 2/14/2014
5
Paddle Electrode Pad Electrode
Department of Biomedical Engineering, SSUET
9
DEFIBRILLATOR ELECTRODES
DEFIBRILLATOR ELECTRODES
Department of Biomedical Engineering, SSUET 10
6. 2/14/2014
6
DEFIBRILLATOR ELECTRODES
11
ï‚—Paddle-type Electrodes have insulated handles
ï‚—Designed to prevent the spread of jell from electrodes to handles for the safety and ease of operator
Department of Biomedical Engineering, SSUET
Automated External Defibrillator Electrode
DEFIBRILLATOR ELECTRODES
12
Department of Biomedical Engineering, SSUET
7. 2/14/2014
7
PRINCIPLE OF DEFIBRILLATION
13
ï‚—Energy storage capacitor is charged at relatively slow rate from AC line
ï‚—Energy stored in capacitor is then delivered at a relatively rapid rate to chest of the patient
Simple arrangement involve the discharge of capacitor energy through the patient’s own resistance
Department of Biomedical Engineering, SSUET
PRINCIPLE OF DEFIBRILLATION
14
Department of Biomedical Engineering, SSUET
8. 2/14/2014
8
PRINCIPLE OF DEFIBRILLATION
15
ï‚—A shorter high-amplitude defibrillation pulse can be obtained by using the capacitive-discharge circuit shown in figure (previous slide)
ï‚—In this case, a half-wave rectifier driven by a step-up transformer is used to charge the capacitor C
ï‚—The voltage to which C is charged is determined by a variable autotransformer in the primary circuit
Department of Biomedical Engineering, SSUET
PRINCIPLE OF DEFIBRILLATION
Department of Biomedical Engineering, SSUET
16
ï‚—A series resistance R limits the charging current to protect the circuit components, and an ac voltmeter across the primary is calibrated to indicate the energy stored in the capacitor
ï‚—The resistor also helps to determine the time necessary to achieve a full charge on the capacitor
ï‚—The clinician discharges the capacitor when the electrodes are firmly in place on the body by shortly changing the switch S from position 1 to position 2
9. 2/14/2014
9
PRINCIPLE OF DEFIBRILLATION
17
ï‚—The capacitor is discharged through the electrodes and the patient's torso, which represent a primarily resistive load, and the inductor L
ï‚—The inductor tends to lengthen the pulse, producing a wave shape of the type shown in figure (previous slide 14)
ï‚—This situation is determined entirely by the resistance between the electrodes, which can vary from patient to patient
Department of Biomedical Engineering, SSUET
PRINCIPLE OF DEFIBRILLATION
18
ï‚—Once the discharge is completed, the switch automatically returns to position 1, and the process can be repeated if necessary
ï‚—With a circuit such as this, 50 to 100 J is required for defibrillation, using electrodes applied directly to the heart
ï‚—When external electrodes are used, energies as high as 400 J may be required
Department of Biomedical Engineering, SSUET
10. 2/14/2014
10
Classes of discharge waveform
ï‚—There are two general classes of waveforms:
a)Mono-phasicwaveform
•Energy delivered in one direction through the patient’s heart
a)Biphasic waveform
•Energy delivered in both direction through the patient’s heart
19
Department of Biomedical Engineering, SSUET
Classes of discharge waveform
Monophasicpulse or waveform
Bi-phasicpulse or waveform
20
Department of Biomedical Engineering, SSUET
11. 2/14/2014
11Classes of discharge waveform
Fig:-Generation of bi-phasicwaveform
21
Department of Biomedical Engineering, SSUET
Classes of discharge waveform
ï‚—The biphasic waveform is preferred over monophasicwaveform to defibrillate
ï‚—A monophasictype, give a high-energy shock, up to 360 to 400 joules due to which increased cardiac injury and in burns the chest around the shock pad sites
ï‚—A biphasic type, give two sequential lower-energy shocks of 120 -200 joules, with each shock moving in an opposite polarity between the pads
22
Department of Biomedical Engineering, SSUET
12. 2/14/2014
12
About Lab
23
ï‚—About Lab files
ï‚¡Do not copy paste word to word from Internet
Do not make a photocopy of your friend’s lab file
ï‚¡Marks will be deducted in both above cases
ï‚¡Final viva will be conducted based upon whatever you written in your lab file, make sure you know everything
Department of Biomedical Engineering, SSUET
CARDIOVERTER
Department of Biomedical Engineering, SSUET
24
ï‚—When an operator applies an electric shock of the magnitude of that from a dc defibrillator to the patient's chest during the T wave of the ECG, there is a strong risk of producing ventricular fibrillation in the patient
ï‚—To avoid this problem, special defibrillators are constructed that have synchronizing circuitry so that the output occurs immediately following an R wave, well before the T wave occurs
ï‚—Figure in the next slide is a block diagram of such a defibrillator, which is known as a cardioverter
13. 2/14/2014
13
CARDIOVERTER
Department of Biomedical Engineering, SSUET
25
CARDIOVERTER
Department of Biomedical Engineering, SSUET
26
ï‚—Basically, the device is a combination of the cardiac monitor and the defibrillator
ï‚—ECG electrodes are placed on the patient in the location that provides the highest R wave with respect to the T wave
ï‚—The signal from these electrodes passes through a switch that is normally closed, connecting the electrodes to an appropriate amplifier
ï‚—The output of the amplifier is displayed on a cardioscopeso that the operator can observe the patient's ECG to see, among other things, whether the cardioversionwas successful-or, in extreme cases, whether it produced more serious arrhythmias
14. 2/14/2014
14
AUTOMATIC EXTERNAL DEFIBRILLATOR
27
Department of Biomedical Engineering, SSUET
AED
ï‚—AED is a portable electronic device that automatically diagnoses the ventricular fibrillation in a patient
ï‚—Automatic refers to the ability to autonomously analyze the patient's condition
ï‚—AEDs require self-adhesive electrodes instead of hand held paddles
ï‚—The AED uses voice prompts, lights and text messages to tell the rescuer what steps have to take next
28
Department of Biomedical Engineering, SSUET
15. 2/14/2014
15
ELECTRODE PLACEMENT OF AED
Anterior electrode pad
Apex electrode pad
Fig. anterior –apex scheme of electrode placement
29
Department of Biomedical Engineering, SSUET
WORKING OF AED
ï‚—Turned on or opened AED
ï‚—AED will instruct the user to:-
ï‚¡Connect the electrodes (pads) to the patient
ï‚¡Avoid touching the patient to avoid false readings by the unit
ï‚¡The AED examine the electrical output from the heart and determine the patient is in a shockablerhythm or not
30
Department of Biomedical Engineering, SSUET
16. 2/14/2014
16
WORKING OF AED
ï‚—When device determined that shock is warranted, it will charge its internal capacitor in preparation to deliver the shock
ï‚—When charged, the device instructs the user to ensure no one is touching the patient and then to press a red button to deliver the shock
ï‚—Many AED units have an 'event memory' which store the ECG of the patient along with details of the time the unit was activated and the number and strength of any shocks delivered
31
Department of Biomedical Engineering, SSUET
32
•Current requirements normally range up to 20 A
•Voltage ranges from 1000V to 6000V
•Time of discharge is kept from 5 to 10 msec
•Current is dependent on the body (chest) resistance
Department of Biomedical Engineering, SSUET
17. 2/14/2014
17
IMPLANTABLE DEFIBRILLATORS
33
ï‚—An implantable cardioverterdefibrillator (ICD) resembles a pacemaker, but its circuitry is similar to that in an AED
ï‚—The battery, capacitor, and electronics are enclosed in a metal case which is implanted under the skin in the chest
The typical size of the case, or ‘‘can’’ is about 50x50x15mm
Department of Biomedical Engineering, SSUET
ICD
34
ï‚—The capacitors in an ICD are slightly smaller than in an AED, but in an ICD the capacitor is charged to a voltage of only approx. 600 V, implying a charge of 0.075 C and an energy of 23 J
ï‚—An ICD delivers about one tenth the energy that an AED does, but in an ICD the shock is delivered through electrodes placed within the heart and is therefore just as effective for defibrillation
ï‚—Tissue impedance for an ICD is at least 50 ohms, implying a discharge time constant of 5 ms
Department of Biomedical Engineering, SSUET
18. 2/14/2014
18
ICD
35
ï‚—Lithium batteries use in ICDs
ï‚—Two batteries in series provide about 6 V
ï‚—Since the capacitor voltage is 600 V, the batteries are used to power a high voltage power supply
They are implanted in the patient’s body, so changing them requires surgery, implying that battery lifetime is important the battery performance begins to decay before its total charge is exhausted
ï‚—Also it must provide power for continuous monitoring of the ECG and other functions, so its observed lifetime is 5 years
Department of Biomedical Engineering, SSUET
ICD
36
ï‚—Another important property of a battery is the time required to charge the capacitor
Typically, the battery takes 10–20 sec to generate a full charge
ï‚—If this time increased significantly, it would delay the delivery of the shock
Department of Biomedical Engineering, SSUET
19. 2/14/2014
19
ICD
37
ï‚—The electrodes and their leads are critical components of an ICD
ï‚—Unlike the electrodes in an AED, ICD electrodes are implanted inside a beating heart and must continue to function there for years
ï‚—Many ICD malfunctions arise because of problems with the leads
ï‚—A typical lead contains three electrodes: one for pacing and sensing and two for defibrillation
Department of Biomedical Engineering, SSUET
ICD
38
ï‚—The ICD recording lead senses the several-millivoltECG signal within the heart
ï‚—Two parameters that the ICD uses to detect abnormal arrhythmias are heart rate and arrhythmia duration
ï‚—The ICDs use sophisticated algorithms to determine from the ECG if an arrhythmia is present, and these algorithms differ between manufacturers
ï‚—Sufficient memory is included in the ICD to store ECGs before, during, and after a shock
Department of Biomedical Engineering, SSUET
20. 2/14/2014
20
Reference
39
•For reference visit:
•Book: Encyclopaedia of Medical Devices and Instrumentation 2nded. J. Webster (Wiley 2006)
•http://www.resuscitationcentral.com/defibrillation/biphasic-waveform/
•http://www.nhlbi.nih.gov/health/health- topics/topics/aed/
•http://www.nhlbi.nih.gov/health/health- topics/topics/icd/
•And various other resources on internet
Department of Biomedical Engineering, SSUET