This document outlines guidelines for mammography room layout and shielding as per AERB (Atomic Energy Regulatory Board). It includes dimensions for equipment such as the column, stand, CEPH arm, and cassette holder. It specifies the entry door should have a lead lining of 2mm and a warning sign with red light that is on during x-ray use. The room should be 2m x 2m with 22cm brick, 15cm concrete, or 2.2mm lead shielding, or an adjusted size according to the table. Contact information is provided for the issuing organization.
1 of 6
Downloaded 18 times
More Related Content
Apex OPG installation Guide as AERB
1. OPG Room Layout n Shielding
Guidelines as per AERB
Issued in public interest by:
Column
Stand
CEPH Arm
CEPH Cassette Holder
(for CR imaging plate)
103cm
50cm
50cm
Approx.3or93cmDepth(standbyposition)
Tube Head
Detector or
Cassette
Assembly
Foldable
CEPH arm
Approx. 3 or 93cm in case of without CEPH
Approx. 3 or 93cm additional space
required for CEPH attachment
Approx. 6' or 186 cm
ENTRY DOOR
with Lead lining of 2mm
NO ENTRY WHEN RED LIGHT IS ON
ENTRY DOOR
should have following
warning sign displaced
with Red Light Bulb
which should be ON
during X-ray machine
is in use / X-ray ON
Room of 2m x2m with 22cm Brick or 15cm concrete or 2.2mm lead
or room size can be changed according to the above table
Apex Medical Systems Pvt. Ltd.
Corporate Office:
A-218, Som Dutt Chambers-I
5-Bhikaji Cama Place
New Delhi: 110066
INDIA
www.apexsystems.in
Email: service@apexsystems.in
Land lines: +91 11 6563 7788, 2617 4069
PK Sharma +91 98110 57761
pksharma@apexsystems.in
Sanjay Chadha +91 98110 56769
sanjaychadha@apexsystems.in
Sanjay Pubral +91 98997 88501
Projection View for layout visualization NOT TO SCALE
Centre of
Patient Table
2. Mammography Room Layout Guidelines
as per AERB
Issued in public interest by:
Apex Medical Systems Pvt. Ltd.
Corporate Office:
A-218, Som Dutt Chambers-I
5-Bhikaji Cama Place
New Delhi: 110066
INDIA
www.apexsystems.in
Email: service@apexsystems.in
Land lines: +91 11 6563 7788, 2617 4069
PK Sharma +91 98110 57761
pksharma@apexsystems.in
Sanjay Chadha +91 98110 56769
sanjaychadha@apexsystems.in
Sanjay Pubral +91 98997 88501
Column
Stand
Projection View for layout visualization NOT TO SCALE
50cm
50cm
Approx.3or93cmDepth(standbyposition)
Tube Head
Detector or
Cassette
Assembly
Approx. 3 or 93cm i
ENTRY DOOR
NO ENTRY WHEN RED LIGHT IS ON
ENTRY DOOR
should have following
warning sign displaced
with Red Light Bulb
which should be ON
during X-ray machine
is in use / X-ray ON
Room of 2m x2m with 1.5cm Brick or 1.0cm concrete or Gypsum Wallboard of 1.5cm
or room size can be changed according to the above table
Centre of
Patient Table