1. The document provides instructions for distilling beer samples to analyze for diacetyl content using the ASBC Method Beer 25A.
2. The process involves adding potassium phosphate buffer and sodium hydroxide to the beer in a distilling flask, then slowly distilling the mixture over 50 minutes to collect 30mL of distillate in a bubbler chamber.
3. The distillate is then treated with reagents and its absorbance is measured spectrophotometrically against a blank to determine the diacetyl concentration in the original beer sample.
1 of 4
More Related Content
Diacetyl
1. UNCLASSIFIED
UNCLASSIFIED
1 March 2005
Page 1
Distillation Setup for ASBC Method Beer 25A
These notes contain the suggested ways to process beer samples for diacetyl analysis. The
basic Apparatus setup is shown in Figure 0.1
1. Place 500 mL mantle on counter
2. Place 500 mL distiling ?ask with gas introduction neck in mantle
3. Connect gas neck to output of ?ow meter, input to bottle
4. Open stopcock. No gas ?ow at this point
Figure 0.1 Apparatus con?guration for distillation of beer
samples for diacetyl analysis. Note receiver (in beaker
on jack stand) is surrounded by ice. Also note
carborundum boiling stones, Antifoam A and carbon
dioxide ?ow meter. (28Jun00 - DSCN0222.pct)
CO2
2. UNCLASSIFIED
UNCLASSIFIED
1 March 2005
Page 2
5. Weigh out 1 gram of monobasic potassium phosphate and
place in bottom of ?ask
6. Add 25 mL of 0.1N NaOH. Swirl to mix
7. Measure out 250 mL of cold, undecarbonated beer into volumetric
?ask. Suck foam out with pipet.
8. Add a few carborundum boiling stones to ?ask.
9. Pour beer in gently. Swirl to mix with buffer.
10. Add about 0.1 mL Antifoam A. The ¡°slick¡± should be 1 cm in diameter
or a little larger.
11. Place distillation column in neck of ?ask. Secure neck of ?ask and col-
umn with clamps on vertical rod of mantle.
12. Place Claisen adapter at top of column. Stopper top output.
13. Place a ring stand to right of mantle.
14. Attach condenser with cooling connections already made to sloping
port on Claisen adapter. Stabilize with clamp on second stand. Orient
with hose connections down so that drip tip will be down.
15. Place jack stand under outlet end of condenser.
16. Place 500 mL beaker on jack stand.
17. Place 0.8 mL of 6% hydroxylamine hydrochloride ( 0.6 gram in 9.4 mL
water) in the bottom of 30 mL bubbler chamber. Add about 2 mL DI
water.
18. Connect bubbler to vacuum adapter, connect vacuum adapter to con-
denser and raise jack stand so that bubbler is supported by bottom of
beaker on jack stand.
19. Check ground glass connections for tightness and turn on . Estab-
lish a ?ow reading of 60 on ?ow meter (144 ml/Min).
20. Wait 5 minutes for air to be swept out of apparatus. Reduce ?ow to 10
(12 mL/min).
21. Set mantle to 10.
KH2PO4
CO2
3. UNCLASSIFIED
UNCLASSIFIED
1 March 2005
Page 3
22. After 13 minutes move heat back to 6. There should be foam on surface
of beer and some re?ux from top of ?ask but none from column yet.
Watch foam situation like a hawk and be prepared to back off heat to 5
or less if there is danger of foam entering the column.
23. After a few minutes, re?ux should begin from column. Alcohol in re?ux
should be enough to collapse the foam. After 18 minutes or so foam
should be under control. Advance heat top 7 when foam is under con-
trol.
24. By 22 minutes or so drops should be appearing at condenser drip tip at
the rate of a bit less than 1 per sec.
25. Collect until 30 mL have accumulated. Note that vacuum adapter tube
displaces a fair amount of liquid so that liquid level will have to be well
above 30 mL mark on bubbler - about up to bottom of ground glass of
adapter. This will take about 50 minutes.
26. Have 100 mL beaker ready. Disconnect bubbler from adapter. Lower
jack stand. When stand is low enough that bubbler can be removed do
so immediately placing beaker under dip tube so as to collect any addi-
tional drops which fall into beaker.
27. Empty bubbler tube into beaker and replace on vacuum adapter. Raise
jack stand and large beaker to support.
28. Grab small beaker with ringstand tube clamp (it will just do it) and
lower into 80C water bath. Be sure bath is full enough to cover liquid in
bottom of beaker.
29. Collect another 5 mL or so of distillate to rinse bubbler. Swirl to collect
any dimethylglyoxime left on walls of tube and add to beaker. Total
time to this point should be just under an hour from when heat was
turned on.
30. Leave in water bath at 80C for 15 minutes.
31. Transfer to boiling water bath until reduced to 15 mL. This should be
about another 15 minutes for total of 1:30 to this point. Cool to room
temp.
32. Transfer quantitatively to 25 mL volumetric ?ask using thistle funnel.
Leave enough space for addition of 3.3 mL additional reagents.
33. Put 25 mL DI water into a volumetric ?ask.
4. UNCLASSIFIED
UNCLASSIFIED
1 March 2005
Page 4
34. To each ?ask add 1.0 mL solution (Reagent c). Allow to stand 5
minutes.
35. Successively add to each ?ask 0.6 ml (Reagent d), 2.5 mL satu-
rated tartrate solution (Reagent c) and 0.2mL (Reagent f).
36. Read diacetyl concentration against blank in 1¡± cell using DR/4000U
User Method 9.
K2HPO4
NH4OH
FeSO4
Figure 0.2 Sample screen from spectrophotometer during
diacetyl assay of a beer. (28jun00 DSCN0223.PICT)