The DG-105 cation conductivity degasser removes dissolved gases from water samples to obtain an accurate measurement of conductivity. It subjects samples to boiling and cooling to degas them before measuring conductivity, which is important for monitoring boiler water. The compact device integrates all necessary components for degasification and conductivity measurement according to industry standards.
Convert to study guideBETA
Transform any presentation into a summarized study guide, highlighting the most important points and key insights.
1 of 3
Download to read offline
More Related Content
Ds degassed
1. DESCRIPTION
The cation conductivity degasser DG-105 is a device designed to obtain the value of the conductivity of a
sample, eliminating the contribution to it of possible gases that may be dissolved. After subjecting the sample
to boiling, releasing any gas can be dissolved, a cooling is performed at 25-30 属 C, and then the treated sample
to a conductivity cell is introduced.
Degassed cationic conductivity is an important parameter for the boiler, most boiler manufacturers have
speci鍖ed that the value of the cation conductivity in the steam must be below a certain limit of conductivity (less
than 0.5 microsecond / cm), to avoid damage to it or the turbine.
APPLICATIONS
Steam-water analysis systems.
FEATURES
According with ASTM D4519
Including all necessary components as a complete unit, just piping, plug-in and ready to use.
Provide with necessary shell protection to prevent operators from hot injury.
Compact wall mount design.
Sample cooler ensures CC and DCC probes receive identical sample temperatures, eliminating errors
(Optional).
Flow indication.
Switch protection No Flow
CE Marking
DEGASSED CATION CONDUCTIVITY
DG 105. SH.1
DEGASSED CATION CONDUCTIVITY
DG 105. SH.1. DG105r00.2017
DEGASSED CATION
CONDUCTIVITY
DG 105
2. www.analyzersmanvia.com
Poligono Ria単o III, Parcela 43 33920 Ria単o, Asturias, Spain
Phone:+34 984204549. Email: analyzer@manvia.com
Set point
A.- Manual
B.- Automatic (atmospheric pressure conpensation)
Sample Cooling
0.- Self cooled
1.- Cooling water
Electrical supply
C.- 120 VAC 60 Hz
D.- 230 VAC 50 Hz
ORDERING INFORMATION DG-105 -- -- --
DEGASSED CATION CONDUCTIVITY
DG 105
SPECIFICATIONS
Max Pressure/Temp Rating:
Sample Flow:
Sample Connection:
Cation Resin:
Cooler Cooling Water Connect. (Only Opt. 1)
Cooling Water Consumption (Only Opt. 1):
Input Power:
Power Consumption:
Plate mounting:
Enclosure Rating:
Dimensions (HxWxD)
4 bar(g) / 50尊C
80 - 250 ml /min.
1/4 卒卒 OD inlet / outlet.
600 cc
1/4 卒卒 NPT(f)
4 l/min
110-120/220-240 VAC, 50/60 Hz.
1.6 KVA
Stainless steel
IP66 / NEMA 4X
900 x 450 x 300 mm
SPARE PARTS
Part #
H1500-120
H-1500-230
RD-105
STH-TXIH-2
MV-PT105
SC-1
Name
Heating Cartidge 120 VAC
Heating Cartidge 230 VAC
Solid State relay 110-230 VAC 50/60 Hz
Turbine
Temperature sensor
Sample cooler
DEGASSED CATION CONDUCTIVITY
DG 105. SH.2
DEGASSED CATION CONDUCTIVITY
DG 105. SH.2 . DG105r00.2017
3. www.analyzersmanvia.com
Poligono Ria単o III, Parcela 43 33920 Ria単o, Asturias, Spain
Phone:+34 984204549. Email: analyzer@manvia.com
DEGASSED CATION CONDUCTIVITY
DG 105
DEGASSED CATION CONDUCTIVITY
DG 105. SH.3. DG105r00.2017
MAIN ELEMENTS
1. - Mounting Plate
2. -Electrical box
3.- Sample Cooler
4.- Turbine
5.- Heater & Thermocouple
DRAWINGS
in mm
Self Cooled Option Cooling Water Option
*Only in Cooling Water
Option