2. Overview
Film-screen radiography
Processing
Intensifying screens and the film cassette
The characteristic curve and sensitivity
Image quality
3. The story so far
We know how X-rays are made in the X-ray tube and
how they interact with the patient
We know how we control the quality and intensity of the
X-ray beam, and hence patient dose, with kVp, mAs,
filtration and distance
We discussed the main descriptors of image quality
Contrast
Spatial Resolution
Noise
Discussed ways to improve contrast by minimising
scatter and using contrast agents
Remember, there is always a balance between patient
dose and image quality fit for the clinical task!
4. Film-Screen Imaging
Traditionally, all X-ray image capture has been
through X-ray film
Film base
Emulsion
Emulsion
Adhesive
layer
Protective
layer
5. Film
Polyester film base gives mechanical strength to
the film does not react to X rays
Emulsion consists of silver halide grains (AgBr)
The image is formed by the reaction of AgBr grains to
X-ray photons
The sensitivity of the film depends on number of grains
Must be evenly distribution
Typically each crystal is about 1 亮m in size
larger grains = more sensitive (contrast),
smaller grain = better resolution
Adhesive layer ensures emulsion stays firmly
attached to base
Protective layer prevents mechanical damage
6. Film
Film is actually much more sensitive to visible
light and UV than it is to X-rays
Hence, use a fluorescent screen to convert X-ray
photons to light photons
Enables lower patient dose!
A latent image is formed upon exposure, which
cannot be seen unless the film undergoes
chemical processing
Mobile silver ions are attracted to electrons liberated
by light photons, forming a speck of silver metal on
the surface
7. Processing
The invisible latent image is
made visible by processing
There are three stages to this
process;
Development
Fixing
Washing
8. Processing
First stage is development:
Film is immersed in an alkaline solution of a reducing
agent (electron donor)
Reduces positive silver ions to metallic grain of silver
(black specks)
Unexposed crystals are unaffected by the developer
bromide ions repel the electron donor molecules
However, given sufficient time, the developer will
penetrate the unexposed crystals
The amount of background fog is dependent upon the
time, strength and temperature of the developer
9. Processing
Second stage is fixing:
If the film is exposed to light after the first stage, the
whole film becomes black
To fix the film, unaffected grains are dissolved by an
acid solution, leaving the X-ray image in the form of
black silver specks
Final stage is washing:
The film is washed in water and dried with hot air
Inadequate washing would result in a brown/yellow
film over time (from excess acid) and smell
10. Processing
Automatic processors use a roller system to
transfer the film through the different solutions
Regular Quality Assurance of the processor is
vital for producing good quality radiographs
Image is then viewed by transmission of light
from a light box with uniform brightness
Dark = lots of X-rays
Light = relatively few X-rays e.g. through bone
11. Production of a Radiograph
Process Time What Happens
1. Manufacture Crystals of a suitable size
are made and suspended
in gelatine
2. Exposure 0.01 10 sec Latent image created
3. Wetting 10 sec Wet film so that
subsequent development
is uniform
4. Development 3 10 min Convert latent image to
silver
5. (Acid) wash 1 min Stop development and
remove excess developer
6. Fixing and hardening 10 30 sec Dissolve out remaining
AgBr and harden gelatine
7. Washing 30 sec Remove products of
developer and fixer
8. Dry 30 sec Remove water
12. Logarithms
A logarithm is an exponent the exponent to
which the base must be raised to produce a given
number
104 = 10x10x10x10 = 10,000
= log1010000 = 4
i.e., 4 is the logarithm of 10000 with base 10
Seen in many applications
Richter earthquake scale
Sound level measurements (decibels = dB)
Optical Densities blackness on film (OD)
Written as log10x or if no base specified in physics
texts as log x it is interpreted as the same
14. Optical Density
Optical Density: the
amount of blackening in
the film
Defined as the log of the
ratio of the intensities of
the incident and
transmitted light
log is used as the eyes
response is logarithmic
15. Optical Density
Optical density can be measured with a
densitometer
From the definition, if 1% of light is transmitted,
D = 2.0
If 10% is transmitted, D = 1.0
The density of an area of interest on a
properly exposure film should be about 1.0
Lung field may be ~2.0
Areas with D>3.0 too dark to see any detail on a
standard light box
16. Contrast
Contrast is the difference in optical densities
Contrast = OD1 OD2
High contrast - e.g. black and white
Low contrast e.g. grey and grey!
17. Intensifying screens
Film is relatively insensitive to X-rays directly
Only about 2% of the X-rays would interact with the
emulsion
Requires unacceptably high doses to give a
diagnostic image
An intensifying screen is a phosphor sheet the
same size as the film, which converts the X-rays
to a pattern of light photons
The intensity of the light is proportional to the
intensity of X-rays
The pattern of light is then captured by the film
One exception is intraoral dental radiography, where
screens are not practical
18. Intensifying screens
Modern intensifying screens use rare earth
materials, which emit light that is matched to the
sensitivity of the film being used
Spectral match between the emission of the screen
and the absorption in the film e.g. blue or green
K-edges clinically relevant (39-61 keV)
Rare earth screens used as they very efficient at
converted absorbed X-ray energy into light
Results in a faster (more sensitive) system
The sensitive emulsion of the film must be in
close contact with the screen
19. Intensifying screens
General radiography film usually double coated
with emulsion on each side of the base
The front screen absorbs ~1/3 of X-rays and
~1/2 light travels forward and is absorbed by
front layer of emulsion
Rear screen absorbs ~1/2 of X-rays transmitted
through the front and exposes the rear emulsion
~2/3 of total X-ray fluence absorbed in screens
Mammography only uses a single screen to
maximise spatial resolution (more on this later)
Screen materials chosen to have no
phosphorescence (delayed fluorescence) to
avoid ghost images
20. The film-cassette
Flat, light tight box with
pressure pads to ensure
film in good contact with
the screen(s) mounted on
the front (and back)
The tube side of the
cassette is low atomic
number material (Z~6) to
minimise attenuation
Rear of cassette often lead
backed to minimise back
scatter (not in mammo)
21. The characteristic curve
Plotting OD against
log exposure gives
the Characteristic
Curve of the X-ray
film
Different types of film
subtle differences
but all basically the
same
Log exposure
Optical
density
Fog
Linear region,
gradient = gamma
Saturation
Solarisation
22. The characteristic curve
Depends on type of film, processing and storage
Fog: Background blackening due to
manufacture and storage (undesirable)
Generally in the range 0.15-0.2
Linear portion: useful part of the curve in which
optical density (blackening) is proportional to the
log of X-ray exposure
The gradient of the linear portion determines
contrast in an image and patient exposures
must lie within this region
Need to match this to the clinical task!
Hence, film suffers from a limited and fixed
dynamic range
23. The characteristic curve
Gradient of linear region =
Gamma, = OD2 OD1
log E2-log E1
Gamma depends on
Emulsion
Size and distribution of
grains
Film developing
Gamma ~ Contrast
Latitude = useful range of
exposures
Linear
region
Latitude
Log exposure
Optical
density
24. The characteristic curve
Gamma and latitude are inversely related
High gamma = low latitude
Wide latitude (low gamma) for chests
High gamma (low latitude) for mammography
At doses above the shoulder region, the curve
flattens off at D~3.5
Saturation, whereby all silver bromide crystals have
been converted to silver
At extremely high exposures density will begin to
fall again due to solarization
Not relevant to radiography
25. Film Speed
Definition: 1 / ExposureB+F+1
Reciprocal of Exposure to cause an OD of 1
above base plus fog
Speed of film = sensitivity = amount of radiation
required to produce a radiograph of standard
density
Speed shifts H-D curve left and right
Fast film requires less radiation (lower patient
dose)
Speed is generally used as a relative term
defined at a certain OD; one film may be faster
than another at a certain point on the curve
26. Factors affecting speed
Size of grains larger means faster
This is the main factor and conflicts with the need for
small crystals to give good image sharpness.
Fast films are grainier but reduce patient dose
Thickness of emulsion
Double layers of emulsion give faster films
Radiosensitisers added
(X-ray energy)
27. Effect of developing conditions
Increasing developer temperature, concentration
or time increases speed at the expense of fog
Developer conditions should be optimised for
maximum gamma, and minimum fog
Automatic processor has temperature controls
and time maintained by roller speed
Concentration is controlled by automatic
replenishment of the chemicals
28. Film-screen sensitivity
Intensification factor
Each X-ray photon generates ~1000 light photons
Just under half of these will reach the film
~100 light photons to create a latent image
Hence, more efficient process
Intensification factor is the ratio of air KERMA to
produce D = 1 for film alone, to that with a screen
Intensification factor typically 30-100
Speed class
Most common descriptor of sensitivity
Speed = 1000/K, where K is air KERMA (in 亮Gy) to
achieve D = 1
Typically 400 speed (K = 2.5 亮Gy)
29. Image quality
Contrast
Contrast in film-screen radiography is due to both
subject contrast, scatter and gamma
Remember, high gamma = high contrast = low
latitude (and vice-versa)
Contrast is fixed for any given film and processing
conditions
Image detail may be lost if contrast is too high as it
may be lost in the saturated or fog regions
Hence, vital to match gamma to the clinical task
Ambient light conditions and viewing box uniformity
may also impact on the subjective contrast presented
to the Radiologist
Use a darkened room, mask off unused areas of lightbox, etc
30. Image quality
Screen-unsharpness
The film-screen system has inherent unsharpness
additional to geometric, motion and absorption
Only partly due to finite size of the emulsion crystals
Most significant effect is due to spread of light from
the point of X-ray absorption in the phosphor, to
detection by the film
Depends on the point in the phosphor where the
interaction occurs
Thicker phosphor layers more sensitive (absorb more
X-rays), but result in more blurring allow lower
patient doses
32. Screen unsharpness
Speed class should be chosen carefully to
match the application
e.g. 400-speed (thick phosphor) for thick sections of
the body (abdo/pelvis),
e.g. 100-150-speed (thin phosphor) for extremeties
(require detail)
Also may have reflective layer on top of
phosphor to increase sensitivity (reflect light
photons back to the film) at the expense of
resolution
Colour dyes to absorb light photons at wider
angles (longer path lengths) at the expense of
sensitivity
33. Screen unsharpness
Crossover light photons from the front screen
may be absorbed by the rear emulsion (and
vice-versa)
Crossover is a significant contributor to overall
unsharpness
Reason for only using one screen in mammography
where resolution is critical
Minimise screen-unsharpness by ensuring good
contact between the screen and film
Poor contact may result from damage to the film
cassette
34. Film-screen in clinical practice
Kilovoltage: Increased kV gives
Increased penetration = lower patient dose
Increased exposure latitude = larger range of tissues
displayed, BUT lower radiographic contrast
Reduction in mAs = shorter exposures = less motion
blur
mAs
Correct mAs must be chosen to ensure the
correct level of blackening on the film avoid
under or overexposing the film
Too much = saturation, too little = thin image
Produce standard protocols that can be adapted for
patient size
35. Exposure Control
For an acceptable image, require a dose at the
image receptor of about 3 亮Gy for film-screen
radiography
This is the exit dose from the patient after
attenuation
Entrance surface dose (ESD) is much higher
than this;
~10 times greater than exit dose for PA chest
~100 times greater for skull
~1000 times greater for AP pelvis
~5000 times greater for lateral lumbar spine
36. Automatic Exposure Control (AEC)
Limited latitude of film makes it difficult to
choose correct mAs skill and experience of
radiographer
Alternative is to use an AEC to terminate the
exposure when enough dose has been delivered
to the film
AEC is a thin radiation detector (ionisation
chamber) behind the grid, but in front of the film
(though in mammo it is behind to avoid imaging
the chamber on the film)
Usually three chambers that can be operated
together or individually
37. Automatic Exposure Control (AEC)
When a predetermined level of radiation is
detected, the exposure terminates
Choice of chambers determined by clinical task
e.g. left and right for lungs in PA chest, but central if
looking at spine
Also has a density control that can increase or
decrease exposure where necessary
AEC limited to exposures in the Bucky system
38. Modern Day
Film is dying out
Across most (but not all) of the country film
is no longer used for General X-ray
imaging
Only mammography (breast imaging),
where very high resolution specialist film is
used
This Trust no longer uses film for
mammography, and is on the verge of being
fully digital
Editor's Notes
#22: Hurter-Driffield Curve, first published 1890, England
Exposure (mAs)
An increase in log exposure of 0.3 = a doubling of the exposure
#24: Large gamma = contrasty film (uniform crystals)
Small gamma = low contrast film (different sized crystals)
Fast film, large grains, likely to be various sizes therefore low contrast
Latitude range of exposures to give the linear response region
High gamma = small film latitude
Low gamma = greater film latitude
Darker films further to right than lighter films but contrast may be the same