WHY?
STATIC MAGNETIC FIELD
1.5 T is 30,000 x the strength of earths
IT IS TOO SRONG.
30,000 Than Earths Magnetic
field
THE MAGNETIC FIELD
IS AT FULL POWER
ALWAYS .IT IS
ALWAYS
ON.
SOO IS IT HAZARDOUS
 STATIC MAGNETIC FIELD
 BIOLOGICAL EFFECTS-MISSILE EFFECT
 MECHANICAL EFFECTS-EFFECT ON FERROMAGNETIC IMPLANTS
 GRADIENT FIELD
 INDUCED CURRENTS (PERIPHERAL NERVOUS SYSTEM)
 AUDITORY DAMAGE
 RF FIELD
 RF POWER DEPOSITION
 BURNS
ZONE
1
THIS REGION INCLUDES ALL AREAS THAT ARE FREELY
ACCESSIBLE TO THE GENERAL PUBLIC.
UNSCREENED MRI PATIENTS ALSO MAY BE ATTENDERS
ALSOZONE
2
THIS AREA IS THE REGION IN WHICH SCREENED MRI PATIENTS AND
PERSONNEL
ZONE
3
ZONE
4
THE MRI ROOM WHERE ONLY SCREENED PATIENTS ARE ALLOWED
AND ON A DIRECT SUPERVISION OF MRI PERSONEL
5 GAUSS LINE
MRI Safety Basics
MRI Safety Basics
INSTRUCTIONS TO THE PATIENT
1. YOU ARE URGED TO USE THE EAR PLUGS OR HEADPHONES THAT WE SUPPLY FOR
USE DURING YOUR MRI EXAMINATION
2. REMOVE ALL MAKE UP AND JEWELRY (E.G., NECKLACES, PINS, RINGS).
3. REMOVE ALL DENTURES, FALSE TEETH, PARTIAL DENTAL PLATES.
4. REMOVE HEARING AIDS.
5. REMOVE EYEGLASSES.
6. REMOVE YOUR WATCH, CELL PHONE, CREDIT AND BANK CARDSAND ALL OTHER
CARDS WITH A MAGNETIC STRIP.
7. USE GOWN, IF PROVIDED, OR REMOVE ALL CLOTHING WITH METAL FASTENERS,
ZIPPERS, ETC.
8.SQUEEZE BULB IS GIVEN TO THE PATIENT FOR EMERGENCY PURPOSES
 H/O MEDICAL DISEASE.
 H/O SURGERY , CANCER TREATMENT (RADIATION SEEDS).
 H/O TRAUMA, WAR INJURY, OCCUPATIONAL HISTORY,
TATTOO
 ALLERGY OR PROBLEM DURING PREVIOUS MRI
 HISTORY OF PREGNANCY AND BREAST FEEDING, LMP
 CLAUSTROPHOBIA AND ANXIETY
Cerebral
aneurysm
clips
Metallic foreign body
in
the eye
Shrapnel,
bullets
(in critical area)
Metallic Ocular implants
Pacemakers Pacing wires
Cochlear implantsNeurostimulators
Implants and metal
Electrical devices
 ONLY METALS THAT ARE MRI COMPATIBLE SUCH AS
TITANIUM ALUMINUM BRASS
 IF YOU ARE NOT SURE IF AN OBJECT IS MRI
SAFEDONT TAKE IT INTO THE ROOM. ASK A
TECHNOLOGIST!!!!!!!!
Usually MRI compatible
Cardiac
 Sternal wires
 Heart valves &
annuloplasty rings
 Coronary stents
 Epicardial wires (cut
short and taped)
 Other vascular
stents, coils and
filters
Other
 Orthopaedic implants
 Contraceptive devices
 Dental implants
MRI PERSONAL SCREENING
 PRE-EMPLOYMENT SCREENING
 REPORTING OF THE EVENTS LIKE TRAUMA, SURGERY ETC.
EMPLOYEE SAFETY
ALL EMPLOYEES MUST BE SCREENED
TO WORK IN A MAGNETIC FIELD
ENVIRONMENT JUST LIKE THE
PATIENTS.
NO EXCEPTIONS.
Device and Object
Screening
 As part of the Zone III site
restriction and equipment
testing and clearing
responsibilities, all sites
should have ready access to
a strong handheld magnet
(1000-Gauss) and/or a
handheld ferromagnetic
detection device.
Few Signs Regarding safety
MRI SAFE MRI Conditional MRI Unsafe
GRADIENT SYSTEM
SAFETY
AUDITORY DAMAGE
 WILL PRODUCE TYPICAL KNOCKING SOUND
INDUCED CURRENTS
(PERIPHERAL NERVOUS SYSTEM)
RESULTS FROM ELECTRO MAGNETIC FIELDS
RESULTS IN HEATING OF HUMAN BODY
o Always apply hearing protection for patients
Auditory damage
o PATIENT MUST BE INFORMED ABOUT THE POSSIBILITY PERIPHERAL
NERVE STIMULATION
o THOSE WHO CLOSE TO THE PATIENT ALSO EXPERIENCE A MILD
PERIPHERAL NERVE STIMULATION
Induced currents
(peripheral nervous system)
RF power deposition
Burns
So..,
It may cause
 Prevent skin to skin contact creating current loops
 Keep sufficient distance between patients skin
and RF coil
 Keep RF Coils straight and parallel to main magnetic
field
RF SAFETY
 In the event of an emergency, you should
first remove the subject from the MRI
scan room
 Stand near the doors to the scan room to
insure no unauthorized emergency
personnel can enter
 NO CODE OR CODE LIKE PROCEDURES
WILL BE RUN IN THE MRI ROOM
HANDLING EMERGENCY SITUATIONS
Emergency stopper button
EMERGENCY STOP BUTTON DOES NOT
QUENCH THE MAGNET THIS TURN OFF MOST
ELECTRICAL POWER IN THE SCANNER ROOM
AND OPERATOR AREA
Stop button
QUENCH
Quench is done in an emergency ,to run the
magnetic field ZERO in order to remove
projectile /patient from the scanner in
An emergency
If these done Cryogens
( Helium and Nitrogen ) BOIL OFF and
DEACTIVATE the magnet
Quench button
FIRE AND
SAFETY
MRI Compatible Fire Extinguisher
REMEMBER THIS MNEMONIC:
MRI =
METAL RESULTS (IN)
INJURY
THANK
YOU

More Related Content

MRI Safety Basics

  • 1. MUHAMMED SWALIH MP FINAL YEAR BSc RADIOLOGY AND IMAGING TECHNOLOGY GANGA INSTITUTE OF HEALTH SCIENCES
  • 3. STATIC MAGNETIC FIELD 1.5 T is 30,000 x the strength of earths IT IS TOO SRONG. 30,000 Than Earths Magnetic field
  • 4. THE MAGNETIC FIELD IS AT FULL POWER ALWAYS .IT IS ALWAYS ON.
  • 5. SOO IS IT HAZARDOUS
  • 6. STATIC MAGNETIC FIELD BIOLOGICAL EFFECTS-MISSILE EFFECT MECHANICAL EFFECTS-EFFECT ON FERROMAGNETIC IMPLANTS GRADIENT FIELD INDUCED CURRENTS (PERIPHERAL NERVOUS SYSTEM) AUDITORY DAMAGE RF FIELD RF POWER DEPOSITION BURNS
  • 7. ZONE 1 THIS REGION INCLUDES ALL AREAS THAT ARE FREELY ACCESSIBLE TO THE GENERAL PUBLIC.
  • 8. UNSCREENED MRI PATIENTS ALSO MAY BE ATTENDERS ALSOZONE 2
  • 9. THIS AREA IS THE REGION IN WHICH SCREENED MRI PATIENTS AND PERSONNEL ZONE 3
  • 10. ZONE 4 THE MRI ROOM WHERE ONLY SCREENED PATIENTS ARE ALLOWED AND ON A DIRECT SUPERVISION OF MRI PERSONEL
  • 14. INSTRUCTIONS TO THE PATIENT 1. YOU ARE URGED TO USE THE EAR PLUGS OR HEADPHONES THAT WE SUPPLY FOR USE DURING YOUR MRI EXAMINATION 2. REMOVE ALL MAKE UP AND JEWELRY (E.G., NECKLACES, PINS, RINGS). 3. REMOVE ALL DENTURES, FALSE TEETH, PARTIAL DENTAL PLATES. 4. REMOVE HEARING AIDS. 5. REMOVE EYEGLASSES. 6. REMOVE YOUR WATCH, CELL PHONE, CREDIT AND BANK CARDSAND ALL OTHER CARDS WITH A MAGNETIC STRIP. 7. USE GOWN, IF PROVIDED, OR REMOVE ALL CLOTHING WITH METAL FASTENERS, ZIPPERS, ETC. 8.SQUEEZE BULB IS GIVEN TO THE PATIENT FOR EMERGENCY PURPOSES
  • 15. H/O MEDICAL DISEASE. H/O SURGERY , CANCER TREATMENT (RADIATION SEEDS). H/O TRAUMA, WAR INJURY, OCCUPATIONAL HISTORY, TATTOO ALLERGY OR PROBLEM DURING PREVIOUS MRI HISTORY OF PREGNANCY AND BREAST FEEDING, LMP CLAUSTROPHOBIA AND ANXIETY
  • 16. Cerebral aneurysm clips Metallic foreign body in the eye Shrapnel, bullets (in critical area) Metallic Ocular implants Pacemakers Pacing wires Cochlear implantsNeurostimulators Implants and metal Electrical devices
  • 17. ONLY METALS THAT ARE MRI COMPATIBLE SUCH AS TITANIUM ALUMINUM BRASS IF YOU ARE NOT SURE IF AN OBJECT IS MRI SAFEDONT TAKE IT INTO THE ROOM. ASK A TECHNOLOGIST!!!!!!!!
  • 18. Usually MRI compatible Cardiac Sternal wires Heart valves & annuloplasty rings Coronary stents Epicardial wires (cut short and taped) Other vascular stents, coils and filters Other Orthopaedic implants Contraceptive devices Dental implants
  • 19. MRI PERSONAL SCREENING PRE-EMPLOYMENT SCREENING REPORTING OF THE EVENTS LIKE TRAUMA, SURGERY ETC.
  • 20. EMPLOYEE SAFETY ALL EMPLOYEES MUST BE SCREENED TO WORK IN A MAGNETIC FIELD ENVIRONMENT JUST LIKE THE PATIENTS. NO EXCEPTIONS.
  • 21. Device and Object Screening As part of the Zone III site restriction and equipment testing and clearing responsibilities, all sites should have ready access to a strong handheld magnet (1000-Gauss) and/or a handheld ferromagnetic detection device.
  • 22. Few Signs Regarding safety MRI SAFE MRI Conditional MRI Unsafe
  • 23. GRADIENT SYSTEM SAFETY AUDITORY DAMAGE WILL PRODUCE TYPICAL KNOCKING SOUND INDUCED CURRENTS (PERIPHERAL NERVOUS SYSTEM) RESULTS FROM ELECTRO MAGNETIC FIELDS RESULTS IN HEATING OF HUMAN BODY
  • 24. o Always apply hearing protection for patients Auditory damage
  • 25. o PATIENT MUST BE INFORMED ABOUT THE POSSIBILITY PERIPHERAL NERVE STIMULATION o THOSE WHO CLOSE TO THE PATIENT ALSO EXPERIENCE A MILD PERIPHERAL NERVE STIMULATION Induced currents (peripheral nervous system)
  • 26. RF power deposition Burns So.., It may cause Prevent skin to skin contact creating current loops Keep sufficient distance between patients skin and RF coil Keep RF Coils straight and parallel to main magnetic field RF SAFETY
  • 27. In the event of an emergency, you should first remove the subject from the MRI scan room Stand near the doors to the scan room to insure no unauthorized emergency personnel can enter NO CODE OR CODE LIKE PROCEDURES WILL BE RUN IN THE MRI ROOM HANDLING EMERGENCY SITUATIONS
  • 28. Emergency stopper button EMERGENCY STOP BUTTON DOES NOT QUENCH THE MAGNET THIS TURN OFF MOST ELECTRICAL POWER IN THE SCANNER ROOM AND OPERATOR AREA Stop button
  • 29. QUENCH Quench is done in an emergency ,to run the magnetic field ZERO in order to remove projectile /patient from the scanner in An emergency If these done Cryogens ( Helium and Nitrogen ) BOIL OFF and DEACTIVATE the magnet
  • 31. FIRE AND SAFETY MRI Compatible Fire Extinguisher
  • 32. REMEMBER THIS MNEMONIC: MRI = METAL RESULTS (IN) INJURY