Equipment Specification: TOC ANALYSER, MICROPROCESSOR
CONTROLLED
Version 2 40.052-1/2
40.052 TOC ANALYSER, MICROPROCESSOR
CONTROLLED
Approval Date: January 2008 Version: 2
Purpose
 The TOC Analyser breaks down all the organically bound carbon molecules
to single carbon unit that can be measured quantitatively.
 To analyse water samples for aggregate carbon content.
Conditions and Requirements
The instrument shall be compatible with the climatic and other environmental
conditions of Indian laboratories.
Specifications
Description  Microprocessor-controlled analyser with facility for
direct measurement of Total Carbon (TC) and Total
Organic Carbon (TOC) in aqueous samples and
suspensions,and purgeable organic carbon (POC) in
samples where purging is possible,having capability
for accepting samples containing suspended solids up
to 500 m size
Method  High temperature catalytic combustion with adjustable
temperature range at least up to 900 oC賊 2 oC drift
 Non-dispersive infrared (NDIR) detector system
Result output
options
 Computer screen
 Printer( if not in-built) and facility for printer port
 ASCII file transferable to floppy
Data processing  Linearisation
 Baseline correction
 Selection of best calibration curve
 Automatic exclusion of anomalous values
 Statistical evaluation, like SD, CV
Operation and
Performance
 Measuring range : 1 mg/L to at least 500 mg/L of
Carbon
 Precision: within 3%
 Detection Limit: at least 500 ppb Carbon
 Linearity: 1% of calibrated range
Power supply  Analyser: 240 volts, 50 Hz  10%
 Auto-sampler: 240 volts, 50 Hz  10%
Equipment Specification: TOC ANALYSER, MICROPROCESSOR
CONTROLLED
Version 2 40.052-2/2
Carrier Gas  Nitrogen or Air of high purity
Sample
introduction
options
 Manual loading through sampling boat for samples
with high suspended matter(> 100 mg/l), and syringe
injection or automatic capillary suction
 Automatic with auto-sampler
Accessories
 Auto-sampler
 Micro-litre syringe 250 L and 2500 L
 Gas cylinder with purified Nitrogen or Air as recommended with regular(2-
stage) and necessary tubings and connectors
 Printer (if not provided as in-built)
 Operation and maintenance manual in English
 Dust cover
 Spares and consumables for two years operation for each of the following:
 Main system
 Auto sampler
 Printer
 Analytical manual including applications for TC ,TOC and POC
measurements
 Instruction and Service manual with one set of the required tools
 Parts catalogue
 Application softwares in English
 Trouble shooting charts

More Related Content

Toc analyser, microprocessor controlled

  • 1. Equipment Specification: TOC ANALYSER, MICROPROCESSOR CONTROLLED Version 2 40.052-1/2 40.052 TOC ANALYSER, MICROPROCESSOR CONTROLLED Approval Date: January 2008 Version: 2 Purpose The TOC Analyser breaks down all the organically bound carbon molecules to single carbon unit that can be measured quantitatively. To analyse water samples for aggregate carbon content. Conditions and Requirements The instrument shall be compatible with the climatic and other environmental conditions of Indian laboratories. Specifications Description Microprocessor-controlled analyser with facility for direct measurement of Total Carbon (TC) and Total Organic Carbon (TOC) in aqueous samples and suspensions,and purgeable organic carbon (POC) in samples where purging is possible,having capability for accepting samples containing suspended solids up to 500 m size Method High temperature catalytic combustion with adjustable temperature range at least up to 900 oC賊 2 oC drift Non-dispersive infrared (NDIR) detector system Result output options Computer screen Printer( if not in-built) and facility for printer port ASCII file transferable to floppy Data processing Linearisation Baseline correction Selection of best calibration curve Automatic exclusion of anomalous values Statistical evaluation, like SD, CV Operation and Performance Measuring range : 1 mg/L to at least 500 mg/L of Carbon Precision: within 3% Detection Limit: at least 500 ppb Carbon Linearity: 1% of calibrated range Power supply Analyser: 240 volts, 50 Hz 10% Auto-sampler: 240 volts, 50 Hz 10%
  • 2. Equipment Specification: TOC ANALYSER, MICROPROCESSOR CONTROLLED Version 2 40.052-2/2 Carrier Gas Nitrogen or Air of high purity Sample introduction options Manual loading through sampling boat for samples with high suspended matter(> 100 mg/l), and syringe injection or automatic capillary suction Automatic with auto-sampler Accessories Auto-sampler Micro-litre syringe 250 L and 2500 L Gas cylinder with purified Nitrogen or Air as recommended with regular(2- stage) and necessary tubings and connectors Printer (if not provided as in-built) Operation and maintenance manual in English Dust cover Spares and consumables for two years operation for each of the following: Main system Auto sampler Printer Analytical manual including applications for TC ,TOC and POC measurements Instruction and Service manual with one set of the required tools Parts catalogue Application softwares in English Trouble shooting charts